5 | Installation
88
D-R 290
The D−SK AE control electronics are electrically connected to the D−SK 290 MA fail-
safe shutter via a 6-core cable.
Do not make the electrical connection to the fail-safe shutter until the measuring
device has been installed!
In addition an F3 air flow sensor which is installed in the purge air infeed to the meas
uring device and which detects any failure of the purge air is connected to the D
−SK AE control electronics.
The 6-core cable (2 metre length supplied) to the fail-safe shutter and the cable to the
air flow sensor can be extended to 50 metres without problem (air flow sensor 3-core
cable with an overall screen).
1
Air flow
2 red colour marking
Fig. 5.21: Air flow sensor
After removing the cover screws (PG 7), install the air flow sensor in the opening
alongside the purge air connection for the measuring device so that the air flow pass
es through the transverse hole in the sensor. The red colour marking is an installation
aid and points away from the device towards the hose (air flow).
D-SK AE electrical connection
The electrical connection to the fail-safe shutter is made using a 20‑way terminal strip
in the D‑SK AE junction box as shown in the connection diagram).
5.5.4
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...