5 | Installation
D-R 290
93
The fail-safe shutter can be closed manually by pressing the push button
“Hand“ (see
the graphic) on the front panel of the control electronics or by parameterising the de
vice settings:
Setting path:
> Specific parameters > D-R 290 configuration > Fail-safe shutter 1 present
or
> Specific parameters > D-R 290 configuration > Fail-safe shutter 2 present
to "0" via D−ESI 100 or D−ISC 100. To activate the shutter function this parameter
must be reset to "1".
The checking for correct operation is performed …
… automatically when the D-SK AE is connected to the measuring head D−R 290. A
brief test is performed every 24 hours.
… manually when the D-SK AE is connected to the measuring head D−R 290 and the
manual test function for the fail-safe shutter is tested.
(see also section
8.2.11 Functional test of the fail-safe shutter [
)
Setting path:
> Functions > D-R 290 Service Functions > Test Functions > Test the fail-safe
shutter 1
> Functions > D-R 290 Service Functions > Test Functions > Test the fail-safe
shutter 2
Commissioning the D-SK AE
Before commissioning the electrical connections must be made.
●
via a 6−core cable between the D‑SK AE control electronics and the
D−SK 290 MA fail-safe shutter (
●
via an 8−core cable between the D−SK AE and the measuring head (
●
via the 3-core cable already installed between the air flow sensor and the D‑SK AE
control electronics (
,
)
Terminals 4 to 9, 16 and 17 must be connected to the measuring head.
Within the measuring head a shutter message must be output:
Fig. 5.24: Control electronics
front panel
5.5.7
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...