7 | Checking / setting parameters with D‑ESI 100
126
D-R 290
Setting path:
Sitemap > »Device (e.g. D‑R 290 ) > >
Save Parameters
For security reasons, parameters that are changed (e.g. specific parameters such as
data for gravimetric calibration) are initially written only to the working memory of the
device. This means these data are currently used or referenced, but are no longer
available after the device has been restarted (i.e. the device then reverts to using the
"old" parameters).
Clicking on the "Save parameters" button causes the data to be transferred to the
read-only memory of the device. This means the parameters are also retained and
continue to be used even the device is restarted.
For the new setting to be used permanently, left click on the "
Save parameters
"
button. (The new setting is not used permanently until it has been saved.)
Limit values (channel #1)… (channel #4)
Each channel can monitor four limit values (two upper limit values and two lower limit
values).
The limit values setting for the measurement channels (channel #1)… (channel #4) is
performed as follows:
Setting path:
Sitemap > »Device (e.g. D‑R 290 ) > >
Common Parameters > Setting
variables > [upper/lower limit value 1/2] >
Set variables for the measurement channels (limit values)
Left click on "Device" in the sitemap.
Left click on for the desired device in the device list.
Under
Common parameters
in the list field, after "
Select group
“ select "
Set
ting variables"
.
Left click on the "Edit" symbol
in the line for which you wish to change the limit
value.
Using the keyboard, correct the entry in the "
Value
" column there.
Repeat the step for each variable whose value you wish to change and for all desired
channels (channel # 1 … # 4)
Step 6
Save parameters
7.5.1
Step 1
Step 2
Step 3
Step 4
Step 5
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...