7 | Checking / setting parameters with D‑ESI 100
D-R 290
141
Fault indication suppression
If a fault is detected by the system, this results in a programmed response. In addition,
the system saves the appropriate (fault) report (see section System messages). The
message is saved in the memory of the message logbook (in the measuring head)
and can provide the maintenance engineer with information for/when carrying out de
vice maintenance, if applicable.
The fault indication suppression designates the time period (in seconds) for which the
system will first pause. The system response is then only initiated if the failure still per
sists after the specified time has expired. This means transient failures can be ignored
without triggering a system response.
Parameters
Default value
Unit
Fault indication suppression
10
Seconds
Table 7.13: Default value for fault indication suppression
Device configuration
The device configuration is device-specific for the D‑R 290 .
Setting path:
Device (
D‑R 290
)
> >
Specific Parameters >
D‑R 290
Configuration
> [PARAMETER]
(Change val
ues manually)
This communicates to the system whether purge air sensors and / or fail-safe shutters
are fitted. After installation and connection of the protective device (at the measuring
head/at the reflector) (see section
5.5.4 D-SK AE electrical connection [
sponding entry must be made at this point in the device configuration.
Fig. 7.13: Device configuration
The device configuration example shown above indicates that only at the measuring
head is a protective device configured (1), whilst the reflector there is no entry (0).
For use of the protective device the value "1" must be entered for
Purge air
sensor x present
and for fail-safe shutter
x installed
.
If the value "0" is entered for
Purge air sensor x present
, the system shows a
"0" from the purge air flow (but does not report an error). Any fail-safe shutter that is
fitted will therefore not be opened.
If the value "0" is entered for fail-safe shutter
x installed
, the status of the fail-
safe shutter is always "4-shutter closed“ (and no error message is output in this case
either).
7.7.3
7.7.4
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...