7 | Checking / setting parameters with D‑ESI 100
142
D-R 290
Measured values
The last measured values from a device are displayed in each case. The selection of
these values is
dependent on the device .
Specially for the D−R 290 the "Values of the last available contamination zero point
and reference point measurement" are presented here.
Setting path:
Device (DR 290) >
>
Specific Parameters >
DR 290 Measured values > [PARAMETER] (Change
values manually)
Fig. 7.14: Measured values
The values of a check measurement not actually in progress show the last check val
ue in each case.
If there is a check measurement actually in progress it shows the continuously upda
ted value.
In the example (illustration above) the measured values relating to contamination, zero
point value and reference point value are shown as "0", since no control cycle has yet
been performed.
If necessary, a contamination warning threshold may be set here.
Note:
If contamination >10% the device outputs an error message in any case. The warning
threshold must therefore be set to a valueless than 10%, if a warning message is
desired earlier than the output by the device of the default error message.
Settings
Setting path:
Device (D‑R 290 ) > >
Specific Parameter >
D‑R 290
Calibration >
7.7.5
Contamination,
zero point value,
reference point value
Contamination
warning threshold
7.7.6
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...