10
Maintenance
Check the safety storage cabinet for any externally visible damage or defects.
Always perform checks:
n
After installation.
n
Before commissioning.
n
After changes.
n
After maintenance.
The safety storage cabinet should also be inspected periodically at the following inter-
vals.
10.1 Inspect daily
Interval Maintenance work
Personnel
Daily
Bottom tray and pull-out trays
n
Check in accordance with regulations governing water legis-
lation
n
Collect and properly dispose of leaked liquids immediately.
Laboratory and warehouse
employees
10.2 Inspect monthly
Interval
Maintenance work
Personnel
Monthly Closing of the doors
n
Open the door and inspect the closure.
Technical specialist employees
Ventilation
n
Check the effectiveness of the ventilation with a woollen thread or
with a smoke tube in the cabinet in front of the left-hand exhaust
air duct at the ventilation grilles.
Technical specialist employees
Seals
n
Check the sealing strips are seated properly in the carcass frame
and on the end faces of the doors.
n
If visible damage is found, replace the sealing strips immediately.
Technical specialist employees
10.3 Inspect annually
Interval
Maintenance work
Personnel
Annually Safety storage cabinet
n
Check of the entire safety storage cabinet
DÜPERTHAL service technicians
Maintenance
Inspect daily
CLASSIC line pure
95
Summary of Contents for 29-200667-006
Page 1: ...Betriebsanleitung Operating instructions CLASSIC line pure Sicherheitsschr nke...
Page 4: ...Erg nzende Anweisungen CLASSIC line pure 4...
Page 6: ...Gesamtinhaltsverzeichnis CLASSIC line pure 6...
Page 7: ...Betriebsanleitung CLASSIC line pure Sicherheitsschr nke...
Page 28: ...Aufbau und Funktion Innenausstattung CLASSIC line pure Lochblecheinsatz CLASSIC line pure 28...
Page 38: ...In Betrieb nehmen Sicherheitsschrank erden CLASSIC line pure 38...
Page 44: ...Betrieb Sicherheitsschrank ohne technische L ftung betreiben CLASSIC line pure 44...
Page 46: ...ffnen des Sicherheitsschranks nach einem Brandfall CLASSIC line pure 46...
Page 50: ...St rungen CLASSIC line pure 50...
Page 52: ...Ersatzteile und Zubeh r CLASSIC line pure 52...
Page 54: ...Entsorgung CLASSIC line pure 54...
Page 56: ...Zertifikate CLASSIC line pure 56...
Page 57: ...Operating instructions CLASSIC line pure Safety storage cabinets...
Page 94: ...Opening the safety storage cabinet after a fire CLASSIC line pure 94...
Page 98: ...Faults CLASSIC line pure 98...
Page 100: ...Spare parts and accessories CLASSIC line pure 100...
Page 102: ...Disposal CLASSIC line pure 102...
Page 104: ...Certificates CLASSIC line pure 104...
Page 105: ...Manuel d utilisation CLASSIC line pure Armoires de s curit...
Page 136: ...Mise en service Mettre l armoire de s curit la terre CLASSIC line pure 136...
Page 144: ...Ouverture de l armoire de s curit apr s un incendie CLASSIC line pure 144...
Page 148: ...D faillances CLASSIC line pure 148...
Page 150: ...Pi ces de rechange et accessoires CLASSIC line pure 150...
Page 152: ...limination CLASSIC line pure 152...
Page 154: ...Certificats CLASSIC line pure 154...