To set or modify channel parameters:
1. Press the
SELECT
button to set or adjust the SET POINT pressure value. Use
the
RAISE
and
LOWER
button to increase or decrease the value shown on the
PRESSURE
display. Pressure range is 1.0 × 10
-5
to 0.1 × 10
-9
in the selected pressure
unit. PRESSURE display, respective pressure unit LED, and SET POINT LED will flash
during adjustment. See
on page 31 for more details.
2. Press the
SELECT
button a second time to adjust the (high) voltage output limit.
Use the
RAISE
and
LOWER
button to increase or decrease the MAX VOLTAGE value
by 0.5 kV that is shown on the HIGH VOLTAGE display. HIGH VOLTAGE display and
MAX VOLTAGE LED will flash during adjustment.
3. Press the
SELECT
button a third time to adjust the current output limit. Use the
RAISE
and
LOWER
button to increase or decrease the MAX CURRENT value shown
on the CURRENT display. CURRENT display, mAMP LED, and MAX CURRENT LED
will flash during adjustment.
4. Press the
SELECT
button a fourth time to set or adjust the pump speed. Use the
RAISE
and
LOWER
button to increase or decrease the PUMP SIZE value by 0.1 L/s
from 0.1 L/s to 9.9 L/s and 1 L/s from 10 L/s to 999 L/s that is shown on the CURRENT
display. CURRENT display and PUMP SIZE LED will flash during adjustment.
5. Press the
SELECT
button a fifth time to set the pressure unit. Use the
RAISE
and
LOWER
button to select the proper PRESSURE UNITS value shown on the PRESSURE
display. PRESSURE display, respective pressure LED, and PRESSURE UNITS LED will
flash during adjustment.
6. Press the
SELECT
button a sixth time to return the unit to standby mode.
All display segments will momentarily illuminate during the setup to standby mode transition.
The unit will return to standby if left unattended during the setup process for approximately 60 s.
Any parameter changes will be saved. If the unit is turned OFF while in setup mode, all modified
parameters will revert to previous values.
A BEEP will be emitted – if the speaker is not
disabled – with each
SELECT
button depression.
NOTE
Since 1976
The selected pressure unit applies to all parameters
and outputs.