Page 13
Changing the Output Polarity
Each channel on the Terranova® 752A can be independently set to output either positive or
negative high voltage.
User should wait at least 15 seconds after turning OFF the
Terranova® 752A before removing the power cord
WARNING!
Figure 4.
Component position for positive voltage output
To change the output polarity:
1. Turn unit OFF via the
POWER
switch
2. Remove power cord and ion pump cable(s)
3. Remove the six #6-32 screws along the edge of the top cover
4. Carefully lift the top cover and disconnect the fan plug
5. Remove top cover
6. Locate and identify the proper HV board for CHANNEL 1 or CHANNEL 2
7. Locate the HV diode. As shown in
, the HV diode is the electrical component
identified by the arrow.