63
6.7
funcon
TARE
In the measuring mode you can carry out a zero point correcon on overpressure sensors
LPC-2
via the
funcon.
TARE
Acvang the
funcon
TARE
Pressing the
key. This sets the current display to 0 and from then on all measure-
TARE
ments are displayed in relaon to the set
TARE
value. The legend
TARE
, displayed
directly above the pressure unit, indicates that this funcon is acve. Press the
key
TARE
again to deacvate the
funcon.
TARE
1. Ba ery level display
2. Indicaon that the
funcon is acve
TARE
3. Current pressure measured value
4. Current unit and type of pressure
5. Bar graph
6.8 The funcon key, "
"
FUNC
6.8.1 Backlighng (
)
LIGHT
Via the
key various coniguraons can be achieved, e.g. the backlighng can be
FUNC
switched on and o f.
f
Acvaon of the backlighng (
) menu item
LIGHT
1
3
2
4
5
Acvated
funcon
TARE
Pressing the
key once* (directly within the measuring mode).
FUNC
* If you are already in a
menu item, press the
key unl the „
“ menu
item appears on the display.
FUNC
FUNC
LIGHT