DRUCK
TEMPERATUR
&
Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt •
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
• www.druck-temperatur.de
GERMANY
DT
www.
-Cal.net
LR
MANUAL
LR
LR
-
-
Cal
Cal
200 +
200-T
LPC
LPC
52
In order to make the electrical connecon for a
reference pressure sensor, bring
LPC-2
together the corresponding cable plug connector to the sensor in line with the orientaon
guide and secure it using a connecon sleeve (screw in the connecon sleeve clockwise
without too much force). To loosen the connecon, the connecon sleeve must be turned
an-clockwise. To disconnect the sensor do not pull the cable, but rather only on the
connector body.
For the connecon to the
, the other end of the cable must also be brought
calibrator
together in line with the orientaon guide and secured using the connecon sleeve (screw
in the connecon sleeve without too much force). To loosen the connecon, the connecon
sleeve must be turned an-clockwise. To disconnect the sensor do not pull the cable,
but rather only on the connector body.
4.5 Power supply
The internal Lithium-Ion ba ery, which can be easily charged with the ba ery charger
supplied with the equipment, serves as the power supply for the instrument.
The instrument is delivered with a charge level of 25 ...
% and should be fully charged
75
once before being used.
The ba ery-level display is shown in the upper right of the screen (see chapter "4.6
Explanaon of display and keypad").
When the mains lead/ba ery charger is connected to the
, the
calibrator
ba ery will be charged, even if the
is switched of .
instrument
f
CAUTION!
The ba ery level during storage or shipping should be between 25 ...
%.
75
•
When the ba ery charger is no longer being used, the mains plug should be disconnected
from the mains socket. Do not leave the ba ery charger connected to the rechargeable
ba ery for longer than one day, since overloading can shorten its service life.
•
Should the rechargeable ba ery sll not be fully charged aer 24 hours, contact the
manufacturer. When not being used, a fully-charged ba ery will lose its charge over me.
•
Extreme temperatures have an adverse ef ect on ba ery charging. As a result, the
f
ba ery may irst need to be either cooled or warmed, as appropriate.
f