DRUCK
TEMPERATUR
&
Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt •
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
• www.druck-temperatur.de
GERMANY
DT
www.
-Cal.net
LR
MANUAL
LR
LR
-
-
Cal
Cal
200 +
200-T
LPC
LPC
47
4. Technical Specificaon
Technical Data
Pressure and Temperature Calibrator
-
200-T
LR
LPC
Cal
(complete measuring chain)
Pressure measurement
1 reference pressure sensor
(changeable without tools)
-2
LPC
1)
Measuring range
bar
0.4 *)
1.6 *)
6 *)
16 *)
40
100
250
600
1000
Overpressure safety
bar
2
10
35
80
80
200
500
1200
1500
Burst pressure
bar
2.4
12
42
96
400
800
1200
2400
3000
Accuracy of the measuring chain
0.025 % FS
2)
Measuring range
bar
1600
2500
4000
5000
6000
Overpressure safety
bar
2300
3500
5000
6000
7000
Burst pressure
bar
4000
6000
8000
10000
11000
Accuracy of the measuring chain
0.1 % FS
2)
Type of pressure
gauge pressure; *) = also as absolute ranges; compound pressure ranges also available:
-1...0 bar, -1...+0.6 bar, -1...+3 bar, -1...+5 bar, -1...+24 bar
Pressure units
bar, mbar, kPa, psi, mmHg, inHg and kg/cm² (depending on the measuring range, selectable)
Acve temperature compensaon
°C
10 ... 40
Temperature measurement
Sensor 295 x 3 mm, with handle
■
Sensor type
Pt100, 4-wire
■
Measuring range
°C
-10 ... +50
■
Resoluon
°C
0.01
■
Accuracy
°C
0.05
Permissible ambient temperature
°C
0 ... 50
Calibraon
2)
Factory calibraon cericate (oponal:
/
kkS calibraon cericate)
DKD DA
Speciicaons
Digital handheld indicator
-
200-T
LR
LPC
Cal
Display
Large graphic display, with backlighng (On/Of selectable)
Display resoluon
up to 6 digits; selectable
Measuring rate (pressure)
5 values/sec
Funcons
Min, Max memory, pressure rate, tare, ofset-correcon for gauge pressure ranges, data logger
funcon
Data logger
- Cyclic data logger: automac recording of up to 1000 values (500 pressure values and 500
temperature values)
- Cycle me: adjustable from 1 ... 3600 seconds
Interface
USB
Permissible
■
Relave humidity
% r. H.
0 ... 85 (non-condensing; at 50 °C)
■
Storage temperature
°C
-20 ... +60
Power supply
■
Current supply
Internal Lithium-Ion rechargeable ba ery (charging me: < 6 h)
■
Ba ery operaon
h
approx. 25
Case
Polyamide 12, membrane keyboard, transparent panels
Ingress protecon
IP 65
Weight
g
approx. 480
CE conformity
■
EMC direcve
2004/108/
,
61326 Emission (Group 1, Class B) and Immunity (portable measurement
EC EN
equipment)
Speciicaons
Reference pressure sensor LPC-2
Pressure connecon
≤ 1000 bar: G ½ B; {various connecon adapters on request}
> 1000 bar: M16 x 1.5 female, with sealing cone
Material
Stainless steel for we ed parts (for measuring ranges > 25 bar ... ≤ 1000 bar addional Elgiloy
®
Internal transmission luid
f
Synthec oil (only for measuring ranges up to 25 bar)
Permissible temperature ranges
■
Medium
°C
-20 ... +80
■
Storage
°C
-40 ... +85
Case
Stainless steel
Electrical connecon
Circular connector, 8-pin
Ingress protecon
IP 65 (with cable connected)
Weight
g
approx. 220
CE conformity
■
Pressure equipment direcve
97/23/
,
> 200 bar; module A, pressure accessory
EC PS
■
EMC direcve
2004/108/
,
61326 Emission (Group 1, Class B) and Immunity (industrial locaons)
EC EN
1) Up to 10 reference sensors
(= pressure ranges) can be used with one instrument
.
LPC
LPC
-2
200-T
LR-
Cal
2) Calibrated at 23 °C and in vercal mounng posion with pressure connecon facing downwards.