13/ 22
Operating instructions for the IFM laser probe model O1D100 (code 3289172)
Image 7
Instructions for calibrating the Laser Probe O1D100
Rotate the wording on the Display by 180°.
1.
Press the
MODE ENTER
key 7 times:
EF.
appears on the Display.
2.
Press the
SET
key.
3.
Press the
MODE ENTER
key 5 times:
diS.
appears on the Display.
4.
Press the
SET
key.
d3.
appears on the Display.
5.
Keep the
SET
button pressed down for 5 sec.
6.
When the wording on the Display no longer flashes, press
SET
once.
7.
rd1.
appears on the Display.
8.
Press
MODE ENTER
once.
9.
Check that the Display wording has rotated by 180°.
Calibrate outlet 1 (OUT 1) operating with window nsP1 (B) & fsP1 (D) (see Calibration table for Laser probe)
1.
Press the
MODE ENTER
key once:
OU1
appears on the Display.
2.
Keep the
SET
button pressed down for 5 sec.
3.
When the wording on the Display no longer flashes, press
SET
twice until
Fno
appears on the Display.
4.
Press the
MODE ENTER
key once:
nsP1
appears on the Display.
5.
Keep the
SET
button pressed down for 5 sec.
6.
When the wording on the Display no longer flashes, press
SET
once.
7.
The value of the height read appears on the Display.
8.
Press the
SET
button until the desired height appears.
9.
Press the
MODE ENTER
button once and the height set is memorised.
10.
Press the
MODE ENTER
key once:
fsP1
appears on the Display.
11.
Repeat the previous points from N° 5 to N° 9.
Calibrate outlet 2 (OUT 2) operating with window nsP2 (A) & fsP2 (C) (see Calibration table for Laser probe)
1.
Press the
MODE ENTER
key once:
OU2
appears on the Display.
2.
Keep the
SET
button pressed down for 5 sec.
3.
When the wording on the Display no longer flashes, press
SET
4 times until
Fno
appears on the Display.
4.
Press the
MODE ENTER
key once:
nsP2
appears on the Display.
5.
Keep the
SET
button pressed down for 5 sec.
6.
When the wording on the Display no longer flashes, press
SET
once.
7.
The value of the height read appears on the Display.
8.
Press the
SET
button until the desired height appears.
9.
Press the
MODE ENTER
button once and the height set is memorised.
10.
Press the
MODE ENTER
key once:
fsP2
appears on the Display.
11.
Repeat the previous points from N° 5 to N° 9.
Applications
The optical distance sensor
measures distances between 0.2 and 10 m.
•
It has a background suppression at > 10..19 m.
•
The measured value is shown in a 10-segment display.
•
According to the set output functions 2 output signals can be generated.
•
O1D100: Certification 21 CFR PART 1040
Operating and display elements
Functions
4.1 Output function hysteresis
The hysteresis keeps the switching state of the output stable if the measured
value varies about the sensing range. Both outputs (OUT1 and OUT2) can be
set as hysteresis function. → 10.2.4 Hysteresis function
4.2 Output function window
The window function enables the monitoring of a defined acceptable range.
Both outputs (OUT1 and OUT2) can be set as window function.
→ 10.2.6 Window function
4.3 Analogue output function
The sensor can also output an analogue signal on output 2 (OUT2) which is
proportional to the distance. → 10.2.11 Scaling of the measuring range
(analogue output).
Electrical connection