Testing
62
Проверка
Checking the Inhalation Valve
1
Tightly pinch the exhalation hose with
your hand. Use peg spanner
16 28 089 of test set R 33 777 if
required.
2
Pump slowly until
3
at least
10 mbar
are indicated on
test unit.
If the system does not reach 10 mbar:
●
Replace inhalation valve or exhalation
valve disc (see “Fitting the Plug-in
Coupling” on page 57).
The low-pressure warning generated by
the Bodyguard during this test is irrele-
vant here.
Checking the Exhalation Valve
1
Connect suction side of the pump to
the test unit.
2
Tightly pinch the inhalation hose with
your hand. Use peg spanner
16 28 089 of test set R 33 777 if
required.
3
Pump slowly until at least
-10 mbar
are indicated on the test unit.
If the system does not reach -10 mbar:
●
Replace exhalation valve or exhala-
tion valve disc (see “Fitting the Plug-
in Coupling” on page 57).
1
5
42
1
7
26.eps
3
2
1
1
552
1
7
26.eps
3
1
2
Проверка клапана вдоха
1
Плотно зажмите рукой шланг
выдоха. При необходимости
используйте штифтовой гаечный
ключ 16 28 089 из испытательного
комплекта R 33 777.
2
Медленно создавайте избыточное
давление, пока
3
испытательное устройство не
покажет минимум
10 мбар
.
Если система не достигает 10 мбар:
●
Замените клапан выдоха или диск
клапана выдоха (см. "Установка
штекерного соединителя" на
стр. 57).
Предупреждение о низком давлении,
которое выдается Bodyguard в ходе
этой проверки, не имеет значения.
Проверка клапана выдоха
1
Подсоедините сторону
всасывания насоса к
испытательному устройству.
2
Плотно зажмите рукой шланг
вдоха. При необходимости
используйте штифтовой гаечный
ключ 16 28 089 из испытательного
комплекта R 33 777.
3
Медленно создавайте
разрежение, пока испытательное
устройство не покажет минимум
-10 мбар
.
Если система не достигает -10 мбар:
●
Замените клапан выдоха или диск
клапана выдоха (см. "Установка
штекерного соединителя" на
стр. 57).