Maintenance and Care
Техническое и сервисное
обслуживание
Maintenance and Care
Note the chapter headed "Maintenance
intervals" on page 73.
Refer to the specific Instructions for Use
of the mask.
Dismantling the PSS BG 4
●
Remove the protective cover (see
“Removing the Protective Cover” on
page 11).
Removing the Oxygen Cylinder
●
Close the cylinder valve.
1
Release anti-vibration device.
2
Open the Velcro closure and pull the
strap out of the buckle.
3
Unscrew the manual connection.
Only by hand!
4
Hold the oxygen cylinder at the
bottom, tip it up and lift it out.
NOTICE
The following sequence should be per-
formed.
DANGER
Ensure that the connections are not
soiled with oil or grease.
To avoid death or serious injury never
allow pressurized oxygen to come into
contact with oil, grease or similar con-
taminants. It could cause either a fire
or an explosion or both.
0502
1
7
26.eps
1
05
1
2
1
7
26.eps
2
3
4
Техническое и
сервисное
обслуживание
Смотри главу ”График технического
обслуживания” на стр. 74.
См. Руководство по эксплуатации
конкретной лицевой маски.
Разборка PSS BG 4
●
Снимите защитную крышку (см.
"Откройте защитную крышку" на
стр. 11).
Как вынуть кислородный
баллон
●
Закройте вентиль баллона.
1
Снимите антивибрационное
устройство.
2
Откройте застежку на липучке и
вытяните ремень из пряжки.
3
Вывинтите маховичок.
Только вручную!
4
Удерживая кислородный баллон за
днище, наклоните его вверх и
выньте из отсека.
УКАЗАНИЕ
Выполните следующие операции в
указанной последовательности.
ОПАСНО
Проверьте, что соединения не
загрязнены маслом или смазкой.
Во избежание гибели или серьезной
травмы никогда не позволяйте
сжатому кислороду контактировать с
маслом, смазкой или подобными
загрязнителями. Это может
привести к пожару, взрыву или
обоим инцидентам.