56
Maintenance
en
4.1
Maintenance intervals
Component
Work to do
Af
ter us
e
Ever
y 6
m
o
n
ths
An
nua
lly
Ever
y 5
y
ear
s
Ever
y 6
y
ear
s
Closed-circuit breathing apparatus Visual inspection, test function and sealing
X
X
1)
X
2)
Clean and disinfect
X
X
Major overhaul
3)
X
CO
2
absorber
(single-use cartridge)
Replace
X
X
4)
CO
2
absorber
(refillable cartridge)
Refill
X
X
4
Compressed oxygen cylinder
Recurring inspection by approved technical authority
X
5)
1) Applies to closed-circuit breathing apparatus that is constantly in use
2) Applies to reserve stocks
3) Only performed by expert personnel. Send the pressure reducer to Dräger for major overhaul.
4) Applies to CO
2
absorbers in closed-circuit breathing apparatus that are stored on vehicles or in the depot, tested in accordance with
chapter 4.5 on page 63 and the connecting piece of which is sealed with the corresponding protective cap.
5) Alternatively: Test in accordance with corresponding national regulations that can allow for differing intervals.
Summary of Contents for PSS BG 4 RP
Page 1: ...Dr ger PSS BG 4 RP i ru 4 en Instructions for Use 40...
Page 2: ......
Page 9: ...9 ru 2 7 2 7 1 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger i i...
Page 12: ...12 ru Dr ger FPS 7000 RP Panorama Nova RP 1 2 2 7 4 PCas APECas 1 2 i i...
Page 20: ...20 ru 4 1 6 5 6 X X1 X2 X X 3 X CO2 X X4 CO2 X X4 X5 1 2 3 Dr ger 4 CO2 4 5 27 5...
Page 35: ...35 ru 5 6 CO2 CO2 7 Dr ger CO2 Dr ger CO2 1 CO2 2 CO2 Dr ger Dr ger CO2 CO2...
Page 38: ...38 ru PC PCas TR...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......