Maintenance
Dräger Polytron 8100
33
6.3
Troubleshooting
6.3.1
Fault reference
Fault number
Cause
Remedy
#001
#003
#004
#005
#011 - #014
#020 - #024
#043
#060
#067
Serious instrument fault,
various causes.
Have the instrument checked by DrägerService
®1
.
#002
#025 - #027
#030 - #034
#041
#042
#044
#050
#052
#081
#085
Serious data error in the instrument,
various causes.
Reset the instrument to factory default settings (see Section 7
on Page 36).
If this fault occurs again:
Have the instrument checked by DrägerService.
#010
4 to 20 mA interface cable not connected.
Disconnect the power supply, connect the 4 to 20 mA interface
and reconnect the power supply.
If operated as a stand-alone instrument without a central
controller, wire pin 3 to pin 4 (see Section 3.5 on Page 10).
#045
Instrument not detecting any sensor.
Check connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
#051
#054
Faulty zero calibration.
Perform zero calibration (see Section 6.2.1 on Page 31).
#055
Faulty span calibration.
Perform span calibration (see Section 6.2.2 on Page 31).
#076
#080 - #084
Instrument fault.
Check electrical connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
#090
Sensor lock function is activated. A sensor
with a different part number has been
connected.
Use a sensor with the same part number or disable the sensor
lock function (see Section 5.10.3 on Page 27).
#105
Instrument fault.
Have the instrument checked by DrägerService.
#137
Instrument fault.
Cycle power.
If this fault occurs again:
Have the instrument checked by DrägerService.
1
DrägerService
®
is a registered trademark of Dräger.
Summary of Contents for Polytron 8100
Page 2: ......
Page 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Page 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Page 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Page 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Page 253: ......