28
Dräger Polytron 8100
Menu
5.10.6 Gas settings
This function sets the gas type, measuring range and units of
measurement.
Not all sensors offer a selection of values.
1. Select
Settings > Sensor > Gas setting
and confirm.
The current measured gas is displayed.
2. Select a measured gas from the list and confirm.
The current unit of measurement is displayed.
3. Select the unit of measurement from the list and confirm.
The current full scale deflection is displayed.
4. Set the full scale deflection and confirm (only for certain
sensors, see sensor data sheet).
The new full scale deflection is displayed.
5. Select
Next
and confirm.
An overview of the new gas settings is displayed.
6. After review: Select
Back to menu
for leaving the function
without changes and confirm with [OK].
7. Select
Confirm
to accept the settings and confirm with
[OK].
Information about units of measurement
Depending on the target gas, the units of measurement can be
switched between:
The instrument automatically converts the sensor signal into
the selected unit of measurement. Units which are not
available for a particular sensor or selected substance will not
be shown.
Information about the full scale deflection (measuring
range) and the analog output of 20 mA.
Some sensors offer a min/max range for the full scale
deflection.
The selected full scale deflection determines the output of the
20 mA signal on the analog interface.
Example:
Required range 0 to 500 ppm CO (e.g. part number 6809605
default 300 ppm, range min/max = 50/1000 ppm).
Select full scale deflection as 500 ppm. The analog output will
be linear between 4 mA = 0 ppm and 20 mA = 500 ppm.
5.10.7 Reset sensor
This function resets all sensor parameters to the factory default
settings see Section 7 on Page 36.
1. Select
Settings > Sensor > Init. channel
and confirm.
2. Select
Confirm
and confirm with [OK].
5.10.8 Calibration interval
After the calibration interval expires, the instrument will issue a
warning that the calibration is past due.
1. Select
Settings > Sensor > Cal. interval
and confirm.
2. Set the calibration interval and confirm.
5.10.9 Set sensor test
This function is only active for Polytron 8100 using the EC
sensor.
This function is only active it the sensor test or diagnostics
dongle is installed.
This function periodically initiates the sensor self-test.
1. Select
Settings > Sensor > Set sensor-test
and confirm.
2. Select
Enable
or
Disable
and confirm.
5.10.10 Fast response
This function is only activated for Polytron 87x0 using the
PIR 7x00 sensor.
5.10.11 Display capture
1. Select
Settings > Sensor > Display capture
and confirm.
The offset can be set in the first window. This value is
output if the measurement value is inside the display
capture range. The lowest value of the capture range,
relative to the capture offset, can be set in the second
window. The highest value of the capture range, relative to
the capture offset, can be set in the third window. All values
are displayed in the 4th window. If one or several values
(capture offset, capture offset - capture low, capture offset
+ capture high) outside the allowed capture range, the
allowed capture range is displayed.
ppb
L/m3
ppm
mL/L
Vol%
mL/m3
%LEL
uL/L
%LIE
uL/m3
%UEG
NOTICE
The calibration interval can be set between 0 and
720 days.
i
i
Summary of Contents for Polytron 8100
Page 2: ......
Page 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Page 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Page 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Page 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Page 253: ......