Technical data
Dräger Polytron 8100
41
10.1.3 DIN EN 50104 for DrägerSensor O
2
LS (6809630)
Sensor Principle
The DrägerSensor O
2
LS (6809630) is an electrochemical
three-electrode sensor for measuring Oxygen (O
2
) in ambient
air.
NOTICE
The sensor cannot be used for Oxygen measurements
in the presence of Helium!
Reaction at electrodes
Measuring electrode
O
2
+ 4 H
+
+ 4 e
–
=> 2 H
2
O
Counter electrode
2 H
2
O => O
2
+ 4 H
+
+ 4 e
–
Environmental parameters
Pressure
20.7 to 38.4 in. Hg (700 to1300 hPa)
Humidity
5 to 95%RH, non-condensing
Temperature
-40 to +60 °C, short term up to +65 °C
Storage temperature
0 to +40 °C
Storage humidity
5-95%RH, non-condensing (relevant
if sensor package is opened)
Influence on environmental parameters
Temperature
In the range -40 to 65 °C
Zero point
< ± 0.3 Vol% O2
Span
< ± 0.3 Vol% O2
Pressure
In the range 700 to 1300 hPa
Zero point deviation
< ± 0.1 Vol% O2
Span: Relative deviation from reading at 1013 hPa: < 2% of
measurement value / 100 hPa
Humidity
In the range 5% r.H to 95% r.H
Zero point
no relevant influence
Span: Relative deviation from reading at 50% r.H: < 2.5% of
measurement value
i
i
Response time
t
0...20
< 14 seconds
t
0...90
< 45 seconds
For measurement range: 0-5 Vol% O2
t
0...20
< 10 seconds
t
0...90
< 27 seconds
For measurement range: 0-25 Vol% O2
In the measurement range 0-5 Vol% the t
0...20
response time
(15 seconds) is higher than the permissible limit of the
EN 50104.
Stabilization time
5 x t
0...90
Calibration
Flow rate
0.5 L/min
Zero gas
Nitrogen (99.9 Vol% N
2
)
Target gas
Oxygen / Nitrogen gas mixture
Recommended calibration interval
9 month
Selectable range:
1 day to 18 months
Warm-up time
Operation
< 20 minutes
Calibration
< 6 hours
Measurement range
0-5...25 Vol% O2, 0-25 Vol%
O2 minimum reading: -1.25 Vol% O2
Summary of Contents for Polytron 8100
Page 2: ......
Page 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Page 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...