Maintenance
40
Instructions for use Polaris Multimedia
Safety checks
.
Safety checks are not a substitute for maintenance,
which includes the preventive replacement of
wearing parts as specified by the manufacturer.
1
Check accompanying documents:
– Instructions for use available
2
Check that the product is in perfect condition:
– Labels complete and legible
– No deformation of the arm system and
loosening of parts
– No paint damage, cracks in plastic parts
– Check the arm system (central axis, swivel
arm, spring arm) for collision damage
(welding seams free of cracks)
3
Check the components and accessories for
completeness according to the instructions for
use.
4
Check electrical safety according to IEC 62353.
5
Check safety features:
– Dimensions of the locking element (at the
spring arm). The display mount and the arm
system must be de-energized.
– Check the position and form of the circlip of
the arm system. The display mount and the
arm system must be de-energized.
Display mount and
arm system
Every 10
years
Electrical safety inspection of
the display mounting system
according to IEC 62353.
Replace components if neces-
sary.
X
Specialized
service per-
sonnel
Component
Interval
Task
Safety
checks
Personnel
responsible
Checks
Interval
Personnel responsible
Safety checks
1)
1) Term applicable in the Federal Republic of Germany, corresponds to the "Periodic safety inspection" in Austria
see tables "Inspec-
tion" and "Preven-
tive maintenance"
see tables "Inspection" and "Preventive
maintenance"
CAUTION
Risk of medical device failure
If the safety checks are not carried out regularly,
the functionality of the medical device may be
jeopardized.
Perform safety checks at the indicated intervals.