Instructions for use Polaris Multimedia
21
Assembly and preparation
Fitting the sterilizable handle
Immediately before an operation, fit the handle
under sterile conditions as follows:
1
Push the handle (A) onto the inner handle. The
handle must engage audibly. If necessary, turn
the handle slightly to the right or left.
2
Check that the handle is securely attached. To
do so, pull in the direction opposite to that used
to attach it.
– To remove the handle, see "Disassembly" on
page 28.
012
020
A
A
A
WARNING
Risk of patient injury
The sterilizable handle can fall down into the
operating field if it is not engaged properly.
The handle must be checked for secure at-
tachment after it has been fitted to the inner
handle.