Maintenance
38
Instructions for use Polaris Multimedia
Inspection
Regular inspections must be carried out according
to the following specifications and at the stated
intervals. Technical documentation is available on
request.
WARNING
Risk of electric shock
Disconnect all electrical connections from the
power supplies before carrying out any work.
Component
Interval
Task
Safety
checks
Personnel
responsible
Display mount and
arm system
Before each
use
Functional and visual inspec-
tion, see page 18.
X
User
Display mount and
arm system
Every 12
months
Check accompanying docu-
ments: instructions for use
available.
X
Service per-
sonnel
Check equipment of the medi-
cal device for completeness
according to the instructions
for use.
X
Perform functional and visual
inspection.
Check for deformations of the
arm system and whether parts
have become loose.
X
Check for paint damage and
cracks in plastic parts.
X
Check whether all plastic parts
are present and in the correct
position.