Reprocessing
28
Instructions for use Polaris Multimedia
Disassembly
Removing the sterilizable handle E
The handle is located at the side of the display
mount and is removed from the inner handle by
pulling.
1
Take hold of the handle section and press the
locking mechanism (A).
2
Keep the locking mechanism pressed and turn
the handle counterclockwise (B) until the
markers (C) line up.
3
Pull the handle (D) down to remove it.
05
0
A
B
051
C
D