Assembly and preparation
20
Instructions for use Polaris Multimedia
Fitting the handle
Fitting the sterilizable handle E
Immediately before an operation, fit the handle
under sterile conditions as follows:
1
Fit the handle (A) on the inner handle so that the
markers (B) line up.
2
Turn the handle (A) clockwise until the locking
mechanism (C) engages audibly.
3
Check that the handle is securely attached.
Turn the handle counterclockwise to do this.
– To remove the handle, see "Disassembly" on
page 28.
046
A
B
047
WARNING
Risk of patient injury
The sterilizable handle can fall down into the
operating field if it is not engaged properly.
The handle must be checked for secure at-
tachment after it has been fitted to the inner
handle.
A
C