![Dräger Evita V300 Supplement To Instructions For Use Download Page 35](http://html.mh-extra.com/html/drager/evita-v300/evita-v300_supplement-to-instructions-for-use_2526753035.webp)
35
Technical data
Essential performance characteristics
Applies to:
– Evita Infinity V500
– Babylog VN500
– Evita V300
The essential performance consists in a controlled
and monitored patient ventilation with user-defined
settings for the monitoring functions
– minimum and maximum tidal volume,
– maximum airway pressure,
– minimum and maximum O
2
concentration in the
breathing gas,
or, if a set limit is exceeded, an appropriate alarm.
Additionally, the integrated monitoring alarms in the
following situations:
– Failure of the external power supply
– Battery discharge
– Failure of the gas supply
Maximum load
Universal holder with standard rail (G93140)
10 kg (22 lbs)
Humidifier holder, can be swiveled (G93111)
5 kg (11 lbs)
Humidifier holder, side rail (8416325)
5 kg (11 lbs)
If a hinged arm is attached to the side rails of the
ventilation unit in addition to the humidifier holder
(8416325), the maximum load of 5 kg (11 lbs) per
side rail must be observed. The humidifier holder
can then only support 4 kg (8.8 lbs).
Operating data (cont.)