Monitoring
18
Monitoring
Flow monitoring
Applies to:
– Evita Infinity V500
– Babylog VN500
– Evita V300
Calibrating Infinity ID flow sensors
Applies to:
– Evita Infinity V500
– Evita V300
Starting the calibration of the Infinity ID
flow
sensor
1
Touch the
Sensors/ Parameters...
button in the
main menu bar.
The device opens the
Sensors/Parameters
dialog
window.
2
Touch the
Flow sensor
tab (A).
3
Touch the
Start
button (B).
4
The device displays information about the
calibration in field (D). Button (E) is preselected.
Confirm with the rotary knob.
The device uses the next inspiratory phase for
calibration of the Infinity ID flow sensor. Short
inspiratory times are extended to
approximately 1 second.
The device displays information about the
calibration in the message field (C).
At the completion of calibration, the
Start
button (B)
turns pale green.
WARNING
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits that
were not removed during reprocessing can
ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor check for
visible damage and soiling, such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particlefree.
11
0
A
B
E
C
Sensors/Parameters
D