![Dräger Evita V300 Supplement To Instructions For Use Download Page 34](http://html.mh-extra.com/html/drager/evita-v300/evita-v300_supplement-to-instructions-for-use_2526753034.webp)
Technical data
34
Operating data
Applies to:
– Evita Infinity V500
– Babylog VN500
– Evita V300
Mains power supply
Inrush current
Approx. 8 to 24 A peak
Approx. 6 to 17 A quasi-RMS
Degree of protection against ingress of liquids and
particles
IP21
Protection against particles with a diameter of more
than 12.5 mm (0.47 in)
Protection against vertically dripping water
Noise emission in accordance with
ISO 80601-2-12:2011 under consideration of
ISO 4871:2009 and ISO 3744:2010
A-class mean surface sound pressure level
(L
PA
) with a radius of 2 m (79 in)
Approx. 33 dB
Approx. 43.5 dB with GS500
Approx. 50 with HFO (Babylog VN500)
Uncertainty (k)
3.5 dB
A-class surface sound pressure level (L
WA
)
Approx. 46 dB
Approx. 57.5 dB with GS500
Approx. 63.5 dB with HFO (Babylog VN500)
Uncertainty (k)
3.5 dB
Weight
Ventilation unit
Approx. 17 kg (37.5 lbs)
Medical Cockpit with holder
Approx. 8 kg (17.6 lbs)
Trolley
Approx. 33 kg (72.8 lbs)
PS500
Approx. 27 kg (59.5 lbs)
GS500
Approx. 10.5 kg (23 lbs)
Nominal weight (weight of ventilation unit and
Medical Cockpit on trolley)
Nom. 58 kg (128 lbs)
Maximum weight (permitted maximum total
weight)
max. 133 kg (293 lbs)