21
Configuration
Setting the priority of the battery alarm
The device offers the following priorities for battery
alarms:
The
Battery activated
alarm message indicating
the changeover to battery operation can be config-
ured as a high- or medium-priority alarm when
Dräger ventilation
is selected.
z
Touch the
Medium
(D) or
High
(E) button and
confirm.
Depending on the setting (IEC/CEI or Dräger venti-
lation), alarm messages have the following priority:
B
IEC/CEI
Priority of the battery alarm in ac-
cordance with IEC 60601-2-12
C
Dräger
ventilation
Priority of the battery alarm ac-
cording to Dräger
Alarm message
Priority
IEC/CEI
Priority
Dräger ventilation
Battery activated
Low-priority alarm message
High-priority or medium-priority alarm
message
Battery low
Medium-priority alarm message
High-priority alarm message
Battery discharged
High-priority alarm message
High-priority alarm message