Technical data
32
Accuracy
±12 % of the measured value or ±12 mL, whichever
is greater, under calibration conditions (1013 mbar
(1013 cmH
2
O), dry gas, 20 °C (68 °F)), 5 % CO
2
,
flow sensor flap closed and no leakage
Exhaled CO
2
per breath
VTCO
2
Range
0 to 550 mL BTPS
Accuracy
±12 %
Applies to:
– Babylog VN500
Device flow (inspiratory, only with HFO)
Accuracy
±10 %
Displayed calculated values
Applies to:
– Evita Infinity V500
Static compliance (determined by Low Flow PV
loop maneuver)
Cstat
0 to 500 mL/mbar (or mL/hPa or mL/cmH
2
O)
Monitoring
Applies to:
– Evita Infinity V500
– Babylog VN500
Sound pressure level L
PA
of alarm signals meas-
ured in accordance with IEC 60601-1-8 and
A1:2012 (3rd edition)
Alarm tone sequence
IEC/CEI
Range for high-priority alarms according to
volume setting
Approx. 55 dB(A) to 72 dB(A)
Range for medium-priority alarms accord-
ing to volume setting
Approx. 52 dB(A) to 69 dB(A)
Range for low-priority alarms according to
volume setting
Approx. 49 dB(A) to 67 dB(A)
Displayed measured values (cont.)