Alarm – Cause – Remedy
!!!
253
Device failure (11)
A failure was detected
during the start-up phase.
Disconnect patient from the
device and continue
ventilation without delay
using another independent
ventilator.
Call DrägerService.
!!!
253
Device failure (12)
A system failure was
detected.
Disconnect patient from the
device and continue
ventilation without delay
using another independent
ventilator.
Call DrägerService.
!!
090
Device failure (13)
The broken wire detection
for the flow sensor is faulty.
Ventilation functions are not
affected.
Call DrägerService.
!!!
237
Device failure (2)
Internal safety system
failure.
Do not use this device for
ventilation therapy.
Call DrägerService.
!!
140
PEEP high
Expiratory valve or
breathing circuit
obstructed.
Check breathing circuit and
expiratory valve.
Check for condensate.
Expiratory resistance
increased.
Check viral/bacterial filter.
Replace it if necessary.
Device failure.
Disconnect patient from the
device and continue
ventilation without delay
using another independent
ventilator.
Call DrägerService.
!!!
140
PEEP low
Applies only to
Babylog VN500.
Measured PEEP is 3 mbar
(3 cmH
2
O) less than set
PEEP.
Check breathing circuit for
tight connections.
Check whether the
expiratory valve is properly
engaged.
Make sure that the tube or
mask is connected
correctly.
Alarm
priority
Alarm message
Cause
Remedy