Configuration
20
Configuration
Applies to:
– Evita Infinity V500
– Babylog VN500
– Evita V300
Information on configuration
The Exchange intervals page has been renamed to
System status
. Information on the exchange inter-
vals can be found at
System setup > System sta-
tus > Exchange intervals
. Additional information
regarding the
System status
page can be found on
page 22.
The
System status
page is password-protected.
Configuring alarm settings
Setting the alarm tone
Setting the minimum alarm volume
Configuring the minimum alarm volume sets the
lower limit of the factory setting for the volume of
the alarm tone (10 % to 100 %). This allows the
setting range to be adjusted to the acoustical situa-
tion at operation site.
Prerequisite: The
System setup > Alarms >
Alarm vol./tone
page is opened.
1
Touch the (A) button.
2
Set the value for the minimum volume by turn-
ing the rotary knob and push to confirm.
WARNING
Missing alarms in loud environments
Alarm situations are not detected.
Set the volume of the alarm tone so that
alarms can be heard.