20
Dräger Alcotest 9510 Standard IR / EC
Making measurement
4
Making measurement
4.1
Requirements
Measurements of breath alcohol concentration with the
Alcotest 9510 may only be carried out by persons trained on
the instrument.
Ensure that the permissible ambient conditions (see Section
12 on Page 27) are met.
Waiting and control time before measurement
Basically, the following has to be considered for breath alcohol
measurement:
Observe a waiting time of 20 minutes after the last
consumption of alcoholic beverages before starting the
measurement.
It must be ensured that during a time period of at least
10 minutes before the measurement no substances are
taken in through the mouth or nose (control time).
Alcohol residues that are still in the mouth may cause
incorrect measuring values.
This may also occur by aromatic drinks (e.g. fruit juices),
alcoholic mouth sprays, medicines and drops, as well as by
burping and vomiting.
Rinsing the mouth with water or non-alcoholic beverages
does not replace the waiting or control time!
4.2
Starting measurement
1. Press the start button to trigger the measuring process.
If the instrument is in energy saving mode (
STANDBY
is
displayed), press the start button again after the warming-
up phase to start the measurement.
Dependent on the instrument configuration the test subject
data are entered prior to the measurement.
The start of the measurement is indicated by the message
PREPARING TEST
.
Complete each entry of corresponding data with
Next
or
Reset
.
4.2.1
Entering data of person being tested
The figure shows the entry via
the virtual keyboard. When
using an external keypad, the
entries must be made
correspondingly via this
keypad. With
it is possible to
change between small and
capital letters.
The entry prompt
NAME OPERATOR
is displayed.
1. Enter the operator name and confirm with
Next
.
The entry prompt
SUBJECT'S FAMILY NAME
is displayed.
2. Enter the test subject name and confirm with
Next
.
The entry prompt
SUBJECT'S FIRST NAME
is displayed.
3. Enter the first name of the test subject and confirm with
Next
.
The entry prompt
SUBJECT'S DATE OF BIRTH
is displayed.
4. Enter the digits for the date of birth (DD), month of birth
(MM) and year of birth (YYYY) and confirm with
Done
.
A summary of the data entered
is displayed. By double click on
the individual entries these can
be edited again.
5. Confirm the entry of test subject data with
Next
. The
measuring process starts immediately after the
confirmation. Scrolling back through or modifying previous
entries no longer possible.
The measuring process is continued.
PURGING
is displayed.
The measuring system is purged with ambient air. Next the
internal reference point is determined,
BLANK CHECK
is
displayed.
Only in the Alcotest 9510 Standard IR / EC this is followed
by a test of the ambient air for alcoholic vapours using the
EC sensor,
ENVIRONMENT CHECK
is displayed.
If interference from the ambient air can be precluded during
this test, the device prompts for providing a breath sample.
The time displayed in the bottom section indicates the time
at which provision of the breath sample must be started.
6. Open the packaging of a new mouthpiece, take the breath
hose out of the storage tray and insert the mouthpiece
without skin contact into the handle at the end of the breath
hose by using the packaging.
NOTICE
If no entry is made within 10 minutes, the message
DATA INPUT TIMEOUT
is displayed and the
instrument returns to
READY
.
Up to 24 letters can be entered for a name. For names
with more letters/characters the last character The
entry is concluded with
Next
.
OPERATOR'S FAMILY NAME
Done
Next
Prev.
Reset
04323875_en.eps
Alcotest 9510
hh:mm
LUEBECK DD.MM.YYYY
1
Q
A
2
W
S
Z
3
E
D
X
4
R
F
C
5
T
G
V
6
Y
H
B
7
U
J
N
8
I
K
M
9
O
L
-
0
P
,
:
+
LUEBECK
i
i
NOTICE
If implausible dates are entered, data acceptance is
refused after selecting the button
Summary
and a
fresh date entry is requested.
Up to this time all previous entries can be recalled on
the display with
Prev.
. The entries can either be
modified by new entries or confirmed with
Done
.
NOTICE
If the external keyboard is being used, the data entry
can also be completed by entering the key
combination [ alt ] + [ s ].
i
i
Save
Reset
02523875_en.eps
Alcotest 9510
hh:mm
LUEBECK DD.MM.YYYY
OPERATOR
Klein
SUBJECT'S FAMILY NAME
Smith
SUBJECT'S FIRST NAME
Paul
SUBJECT'S DATE OF BIRTH
18.09.1956
LUEBECK
i
i