08.08.2018 / Bak We page 25/26
Table 5 (2 of 2)
l)
Direct coupling shall be used..
m)
r.m.s., before modulation is applied.
n)
The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to
13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHZ to 40,70 MHZ. The amateur radio bands between 0,15 MHz and 80
MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHZ, 10,1 MHz to 10,15 MHz, 14MHz to
14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHZ, 24,89 MHz to 24,99 MHz, 28,0 MHZ to 29,7 MHz and 50 MHz to
54,0 MHZ.
o)
Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A / phase and ME EQUIP-
MENT and ME SYSTEMS with RATED input current greater than 16 A / phase.
p)
Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A / phase.
q)
At some phase angles, applying this test to ME EQUIPMENT with transformer mains power input might cause an overcurrent
protection device to open. This can occur due the magnetic flux saturation of the transformer core after the voltage dip. If this
occurs, the ME EQUIPMENT or ME SYSTEM shall provide BASIC SAFTEY during and after the test.
r)
For ME EQUIPMENT and ME SYSTEMS that have multiple voltage settings or auto ranging voltage capability, the test shall be
performed at the minimum and maximum RATED input voltage. ME EQUIPEMT and ME SYSTEMS with a RATED input voltage
range of less than 25 % of the highest RATED input voltage shall be tested at one RATED input within the range. See table 1
Note c) for examples calculations.