Table 9 – Test specifications for enclosure port immunity to
RF wireless communications equipment
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0,3
27
450
430 - 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz deviation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0,3
28
870
930
1720
1700 – 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0,3
28
1845
1970
2450
2400 – 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0,3
28
5240
5100 – 5800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2
0,3
9
5500
5785
NOTE If necessary to achieve the immunity test level, the distance between the transmitting antenna and the ME equipment or ME
system may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a )
For some services, only the uplink frequencies are included.
b )
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c )
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.