7.6 General remarks
When using more than one light at the same time (light combinations), due to the overlapping of
the light fields of different lights, the total radiation intensity can exceed the value of 1000 W/m
2
. This
means a risk of higher heat development in the wound field.
When using more than one light at the same time (light combinations), due to the light fields overlapping
of different lights the maximum permissible values for UV-radiation (< 400 nm) of 10 W/m² can be ex-
ceeded.
The test certificate for the electrical safety test can be requested when needed. Please
provide the serial number of the respective light.
In case of a collective wiring of further lights or devices at installation, chapter 16 of the European
standard EN 60601-1:2013 must be applied and eventually it has to be checked if the requirements
are met.
The light must be tested according to EN 62353 at commissioning.
The polarity is very important for the installation of the light. In case the light does not
function after installation, the polarity must be checked at the secondary side of the power sup-
ply.
8.
CE-mark
The products Mach LED 150F / 150 / 150FP comply with the standards 93/42/EEC for
medical products of the European Community’
s Council.
Dr. Mach applies the standard EN 60601-2-41.
Dr. Mach GmbH is certified according to EN ISO 13485:2012 + AC:2012.
9.
Disposal
The light doesn’
t contain any dangerous goods.
The components of the light should be properly disposed at the end of its shelf-life.
Make sure, that the materials are carefully separated.
The electrical conducting boards should be submitted to an appropriate recycling
proceeding.
The rest of the components should be disposed according to the contained materials.