Mach LED 150F / 150 / 150FP
59320002
Edition 12
08.08.2018 / Bak We page 21/26
11.
Electromagnetic compatibility
The Dr. Mach OT- and examination lights are subject to special preventive measures regarding the
electromagnetic compatibility and must be installed according to the EMC-instructions mentioned in the
accompanying documents
The function of the OT- and examination lights can be affected by portable and mobile HF-communication
or other HF-devices.
If the essential performance should be lost or degraded due to electromagnetic disturbances it is to be
expected that the provisioning of the illumination, for a short time, is not ensured.
To ensure the basic safety and essential performance of the Mach LED 150/150F/150FP regarding elec-
tromagnetic disturbances during the expected service life, the maintenance has to be performed within
the defined interval and according to the instructions in the technical service manual. Only spare parts
specified by the legal manufacturer may be installed.
The Examination light Mach LED 150/150F/150FP has been tested under Professional healthcare
facility environment.
The use of other equipment than the equipment mentioned in chapter 2. leads to an increased
emission or to a reduced interference resistance of the device.
For the intended use of the OT-light MACH LED 150/150F/150FP it is required that the light MACH
LED 150/150F/150FP is not mounted immediate and near to other devices or mounted together
with other devices. If operating the light is obligatory near other devices or together with other
devices, the functions of the light MACH LED 150/150F/150FP must be observed.
Portable RF communications equipment (including peripherals such as antenna cables and exter-
nal antennas) should be used no closer than 30 cm (12 inches) to any part of the Mach LED
150/150F/150FP, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.