Table 2 – Emission limits per environment
Phenomenon
Professional healthcare facility
environment
a)
Home healthcare environment
a)
Conducted and radiated RF emissions
Class B; Group 1 (according CISPR 11)
CISPR 11
c), d)
Harmonic distortion
See IEC 61000-3-2
b)
See IEC 61000-3-2
Voltage fluctuations
and flicker
See IEC 61000-3-3
b)
See IEC 61000-3-3
a)
See 8.9 of IEC 60601-1-2:2014 for information about the environments of intended use.
b)
This test is not applicable in this environment unless the ME EQUIMENT and ME SYSTEMS used there will be connected to
the public mains network and the power input is otherwise within the scope of the Basic EMC standard.
c)
ME EQUIMENT and ME SYSTEMS intended for use in aircraft shall meet the RF emissions requirements of
ISO 7137. The conducted RF emissions test is applicable only to ME EQUIMENT and ME systems that are intended to be
connected to aircraft power. ISO 7137 is identical to RTCA DO-160C:1989 and EUROCAE ED-14C:1989. The latest editions
are RTCA DO-160G:2010 and EUROCAE ED-14G:2011. Therefore, use of
Section 21 (and category M) of a more recent edition, e.g. [39] or [40], should be considered.
d)
Standards applicable to other modes or EM environments of transportation for which use is intended shall apply. Examples
of standards that might be applicable include CISPR 25 and ISO 7637-2.