Fixture Pressure too high
A motor alarm has
occurred. Check
machine for obstruc-
tions. If fault persist,
consult the User’s
Manual, Press the OK
button to clear the fault
Motor Speed Calibration
required
Micro-G® Control Box
NOT on Production
Load Cell Calibration
required
Motor speed too slow
Lost Communication with
Micro-G® Control Box,
Check Control Box and
Network
Motor speed to fast
Lost Communication with
DataLink, Check
DataLink and Network
pressure detected by the polishing machine load cell
varies on the high side by more than the allowed
tolerance from the set point
platen speed detected by the polishing machine varies
on the low side by more than the allowed tolerance
from the set point
if the Micro-G® system is enabled in “User Rights and
Settings” and the polishing machine loses its ethernet
connection to the system the alarm will appear
platen speed detected by the polishing machine varies
on the high side by more than the allowed tolerance
from the set point
if the motor driver electronics for the polishing
machine platen has a fault due to excessive loads or
bad connections the motor alarm will appear
when reloading QP+ program, sometimes this value
gets cleared, Re-run Motor Calibration
when reloading QP+ program, sometimes this value
gets cleared, Re-run Load Cell Calibration
if the DE DataLink® system is enabled in “User Rights
and Settings” and the polishing machine loses its
ethernet connection to the system the alarm will
appear
4
5
6
7
8
9
10
11
12
43
ALARM MESSAGE
CAUSE
Fig. 15
Micro-G® Control box needs to be on its Production
screen