Touching any of the numeric displays for any of the channels will bring up the numeric entry keypad. The
value entered must be between 0.5 and 2.000 for each channel.
5.2.6 Set & Test Outputs
Touching the ‘SET/TEST OUTPUTS’ button on the setup screen brings up the screen shown in figure 5.11.
This allows the user to test relays and analog outputs.
Two other features on this screen:
1.
Require a password to exit analysis
–
If enabled the level 1 password must be used for anyone to exit
the analysis screen. When disabled a simple “Are you sure?” window
appears.
2.
Idle Timeout
–
Set from 1-60 minutes to trigger a fault when the machine is left in idle. Setting this to
0 minutes disables the fault.
Figure 5.11
NOTE
:
This screen requires the Administrator password for access
–
See section
2.4
CAUTION:
Activating these outputs will trigger any connected alarms
Output Relay Test
Alarm Relay Test