Each time the
‘
OPEN/CLOSE GATE
’
button is touched, the gate will open or close appropriately. If the gate is
currently closed, touching the button will open the gate. If the gate is currently open* touching the button will
close the gate.
To Exit the screen optics verification is r
equired. When the ‘Verify Optics’, button is touched the machine will
verify the tape is aligned correctly and the optics are calibrated appropriately. Du
ring this time a ‘ONE
MOMENT PLEASE’ window will appear. When verification is complete one of the t
wo screens in figure 5.8 will
appear.
Figure 5.8
If successful touch the “OK” button to return to the setup menu. If unsuccessful verify that the tape i
s
installed and aligned correctly by using the ‘OPEN/CLOSE GATE’ button as needed. Once you have ve
rified
that the tape is installed correctly you can touch the ‘YES’ button to recalibrate the optics for the new tape or
press ‘NO’ to return to the setup m
enu.
5.2.4 Flow Adjustment
Touching the ‘ADJUST FLOW’ button on the setup screen brings up the screen shown in figure 5.9 along with
advancing the tape and turning the pump on. Use the corresponding flow adjustment knobs located on the
front panel to adjust the flow so that each level is as close as possible to the black line in the middle of the
green section.
NOTE
:
If the gate is not completely open touching the button will open the gate to
the proper open position. Touch the button again to close.
IMPORTANT
:
If you choose ‘NO’ when asked to calibrate the optics as shown to
the right of figure 5.8 the CL4/CL8 may not function properly. Contact DOD
Technologies for more information.