26
DC-ITD-CL4andCL8MAN01.A
www.dodtec.com
NOV 2018
815-788-5200
CL4 & CL8 Manual
Figure 5.2
Figure 5.3
If the operator touches the ‘SETUP MODE’ button before the timer reaches 0 the Setu
p screen appears
(section 5.2) otherwise after a timeout the system will start analysis (section 5.4)
5.2 Setup / Main Menu
The setup menu is accessed either by touching the ‘SETUP MODE’ button on power up (section 5.1) or by
touching ‘SETUP’ from the ana
lysis screen (section 5.5). Figure 5.4 below shows the setup screen and
explains the various buttons.
Figure 5.4
View/Reset tape usage counter.
(Section 5.2.c)
Set system Date & Time (section
5.2.b)
Touch to reset all
active faults &
alarms.
Touch to Start Analysis (Section 5.5)
Touch to
display
version and
configuration
information.
Service password
required
Flow adjustment
–
(section 5.2.d)
Gas selection
and alarm level
settings
(Section 5.2.a)
Adjust Calibration
Factors
(Section 5.2.e)
Set latching outputs and
test outputs.
(Section 5.2.f)
View alarms, system
events, conc. Log,
and other files on
USB Memory Stick.
(Section 5.2.h)