G.3 Open the door and take out the sterilized items.
Warning:
Check the Pressure Gauge is reading ZERO before opening the door.
Warning:
Beware of steam when opening door after a sterilization cycle.
Warning:
Be careful when removing the sterilized items as the metal surfaces might
still be hot. Always wear suitable hand protection to remove the box or use
the appropriate aids (tray holder) to lift the trays.
Warning:
If using the sterilizer continuously, it’s required to have a 20-minute interval
between each sterilization cycle to allow the unit to cool.
Caution:
After the Liquid program, please take out the sterilized items, and then re-run
the “Normal” program to let the water inside chamber return to the water
reservoir.
6.5 Emergency Stop
Press the Vacuum/Pressure Release Button
3 seconds during a cycle to stop the
program and release the pressure/vacuum. The sterilizer will sound to alert. Please wait till the
pressure gauge is reading ZERO, and then the Reset Button
can be pressed to reset the
program.
Warning:
The Vacuum/Pressure Release Button can only been pressed when there’s an
unusual event or emergency. The sterility of the sterilized items should be verified
again.
Warning:
Disposal of the items which is sterilized by unfinished cycle should be in
accordance with the local laws. Do not handle them as general waste.