19
G.2 Press “STERI ITEM”
to select unwrapped instruments or wrapped items. The
corresponding parameters will be:
Sterilization Temperature
121°C
Sterilization Temperature
135°C
Unwrapped
22 Mins
4 Mins
Wrapped
30 Mins
15 Mins
G.3 Press Dry Timer Button
to select a 30-minute drying time. (*If drying is required.)
G.4 Press the START Button
and the sterilizer will automatically run through the
selected program. The current progress of the sterilization cycle is indicated by the
illuminated LED on the Process Status Indicator. (Figure 16)
Figure 16
On completion, the buzzer will sound (long beep). The program has finished when the
buzzer STOPS and the COMPLETE indicator light is illuminated on the Process Status
Indicator.
Warning:
If the COMPLETE indicator light was not lit, the cycle has failed and should
be re-run.
G.5 Open the door and take out the sterilized items. Check the status of the indicators. If failed,
repeat the cycle. Consult with the qualified technician for calibration if necessary. Please
refer to “8. Troubleshooting”.
Warning:
Check the Pressure Gauge is reading ZERO before opening the door.
Warning:
Beware of steam when opening door after a sterilization cycle.
Warning:
Be careful when removing the sterilized items as the metal surfaces might
still be hot. Always wear suitable hand protection to remove the box or use
the appropriate aids (tray holder) to lift the trays.