21
G. How to set the Liquid program
Figure 17
Caution:
After the sterilization phase finished, the unit will take 15 minutes to exhaust
spontaneously. Then the COMPLETE indicator light is illuminated on the
Process Status Indicator.
G.1 Press
to start “Liquid Program”. The corresponding sterilization time will be:
Sterilization Time
40 Minutes
Sterilization Temperature
121°C
Dry Time
N/A
G.2 Press the START Button
and the sterilizer will automatically run through the selected
program. The current progress of the sterilization cycle is indicated by the illuminated LED
on the Process Status Indicator. (Figure 18)
Note: The DRY phase won’t be performed.
Figure 18
On completion, the buzzer will sound (long beep). The program has finished when the buzzer
STOPS and the COMPLETE indicator light is illuminated on the Process Status Indicator.
Warning:
If the COMPLETE indicator light was not lit, the cycle has failed and should
be re-run.