20
Warning:
If using the sterilizer continuously, it’s required to have a 20-minute interval
between each sterilization cycle to allow the unit to cool.
6.3 Re-Dry Program
Press Re-Dry
button and the indicator light will illuminate. Then press START button
to run a 10-minute re-dry program. During the re-dry process, the DRY indicator light will
flash on the Process Status Indicator.
On completion, the buzzer will sound (long beep). The program has finished when the buzzer
STOPS and the COMPLETE indicator light is illuminated on the Process Status Indicator.
6.4 Liquid Program (Optional)
A. Follow “4. Installation” on Page 7 to finish installation first.
B. Follow “4.2 Install the sterilizer” on Page 7 to make sure the water inside reservoir is
sufficient.
C. Check the Pressure Gauge is reading ZERO, and then open the door by turning the door
knob 90° counterclockwise.
D. Place the items to be sterilized into the container as required.
Warning:
Refer to “9. Specifications” for the maximum permissible load. Failure to
follow these instructions may cause the sterilizer to malfunction and result in
an unsuccessful sterilization cycle.
E. Close the door and turn the knob clockwise 90° to lock.
F. Press the “POWER” switch to ON “I” position, the power indicator light should illuminate
with a short beep. The “DOOR” indicator light should be off.