18
6.2 Normal Sterilization Program
A. Follow “4. Installation” on Page 7 to finish installation first.
B. Follow “4.2 Install the sterilizer” on Page 7 to make sure the water inside reservoir is
sufficient.
C. Check the Pressure Gauge is reading ZERO, and then open the door by turning the door
knob 90° counterclockwise.
D. Place the items to be sterilized and the sterilization indicator strips into the box as required.
Remember to open both side windows before placing the box into the sterilizer as Figure
14.
Figure 14
Caution:
Before loading, ensure instruments are cleaned and rinsed.
Warning:
Refer to “9. Specifications” for the maximum permissible load. Failure to
follow these instructions may cause the sterilizer to malfunction and result in
an unsuccessful sterilization cycle.
E. Close the door and turn the knob clockwise 90° to lock.
F. Press the “POWER” switch to ON “I” position, the power indicator light should illuminate
with a short beep. The “DOOR” indicator light should be off.
G. How to set the normal program:
Figure 15
G.1 Press “STERI TEMP”
to select 121°C or 135°C
The cover must be put on during
sterilization box.
Open both side windows before
placing the box into the sterilizer.