17
6.1.2 Operation Overview (Liquid Program) *Optional
*If dry timer is selected.
Press “START” Button
Add Water and Heat
Heat to Selected Sterilization Temperature ● Sterilization Starts
Sterilization Finishes ● Exhaust Starts ● Buzzer Sounds
Select Liquid Program
Open the Door Slightly*
Drying Starts (30 minutes)*
Put the sterilization items and close door
Sterilization Finishes ●
Exhaust Starts
spontaneously (slowly)
Liquid Program
Select Sterilization Temperature
Select Sterilization Time
Select Dry/No Dry
Normal
Sterilization
Normal
Sterilization
Liquid Program
Buzzer Sounds and
COMPLETE indicator light will
illuminate. Unit goes to STANDBY mode.
If the COMPLETE indicator light did not illuminate, repeat
the sterilization cycle.