14
5.2.3 Control Panel (Re-Dry Program)
Figure 12 – Control Panel (Re-Dry Program)
Low Water Indicator Light
Indicates the water level is
below the minimum level or the
program has been unable to
add sufficient water after 5
minutes.
Process Status Indicator
Individual LED light to
indicate the current status of
the sterilization program.
Power
Add Water
Heating
Sterilization
Exhaust – pressure
release and water
drain
Dry
Complete
Visual Alarm Indicator Light
Indicates over-temperature or
over-pressure in the chamber
Emergency Stop Button
Stop program and release
pressure/vacuum.
Power Switch
ON/OFF switch and power
indicator (Illuminates when
ON)
Sterilization Temperature
Button
Select the required sterilization
temperature of 121°C or 135°C.
The selected temperature light
will illuminate.
Sterilization Item Button
Select between unwrapped or
wrapped items. The selected
item light will illuminate.
Dry Timer Button
Select a 30-minute drying time.
The light will illuminate when
selected.
Start Button
Start the selected Program.
Door Open Indicator Light
Indicates the chamber door
is open or not closed fully.
Re-Dry Button
Select Re-Dry Program for
10 minutes.
The light will illuminate
when selected.
Reset Button
Reset program while
running.