Common Issue
1. No Power
1.1 Check to see if power supply is in working order by swapping with a working one.
1.2 Remove top cover and internal cover
.
7.3 Check the voltage.
7.4 Make sure that voltage reads 24V (Scanner must turned on).
7.5 If not, disconnect the power cable on the main board
7.6 Make sure that voltage reads 24V
7.7 If not, replace power board or power switch.
- Refer to the Service Manual p.33 for power board replacement