Basic configuration of CR (=Computed Radiography) system
• Developed and used a , called
“Imaging Plate”
- The function of
“
I
maging Plate”
1) Imaging plate is a flexible plate of 1mm or less thickness coated with fine photostimulable phosphorus
crystal.
2) Temporarily stores the X-ray energy image in portion to the X-ray intensity it receives.
3) The stored energy is released as visible light when a laser in the CR Reader scans the IP.
This phenomenon is generally called “photostimulable luminescence” (PSL)
4) The image on the imaging plate is erased when exposing excessive light.
• Emit the blue light by laser beam scanning.
• The image reader ( = FireCR flash ) converts the latent image on the imaging plate into digital signal.
Phosphor
Base Support
X-ray
Laser beam scanning
White light
E
x
p
o
su
re
Reading
E
rasin
g
Re-use
Summary of Contents for FireCR
Page 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Page 5: ...for User manual download for Quantor software download...
Page 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Page 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Page 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Page 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Page 20: ...Main Board Core Board...
Page 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Page 41: ...2 Disassembly Top Cover 1 Pull back cover...
Page 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Page 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Page 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Page 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...
Page 74: ......