Basic configuration of CR (=Computed Radiography) system
1. X-ray image is stored on IP.
2. It is scanned with a 650 nanometer (nm) laser beam to produce PSL radiation corresponding to the absorbed
x-ray energy.
3. The luminescence radiation stimulated by laser scanning is collected through a light guide into photomuliplier
tube(PMT) which converts it into electric signals.
4. The signal output from the detector is an analog signal and the output from the analog to digital convertor is
time-series digital signal that is being processed.
5.The residual image is erased when exposing excessive light.
Summary of Contents for FireCR
Page 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Page 5: ...for User manual download for Quantor software download...
Page 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Page 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Page 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Page 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Page 20: ...Main Board Core Board...
Page 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Page 41: ...2 Disassembly Top Cover 1 Pull back cover...
Page 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Page 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Page 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Page 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...